Agilent Technologies X Series Manual page 424

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Input/Output
Preset
State Saved
Range
Initial S/W Revision
Q Skew
Sets the skew factor for the Q channel. The skew will shift the channel's data in time. Use this to
compensate for differences in the electrical lengths of the input paths due to cabling and probes.
Key Path
Remote Command
Example
Preset
State Saved
Range
Initial S/W Revision
Q Probe
Accesses the probe setup parameters for the Q channel.
Key Path
State Saved
Readback Text
Initial S/W Revision
I/Q Probe Setup
The set of I/Q probe setup parameters will change based on the type of probe that is sensed. All probe
types have the Attenuation parameter, and all probe types can be calibrated. The remaining parameters
are only available for some probe types and will not be shown when not available. The probe type is
determined by and reported for only for the I and Q ports, never the I-bar or Q-bar ports. The menu title
424
LOW
On
This is unaffected by Preset but is set to the default value on a "Restore
Input/Output Defaults" or "Restore System Defaults->All"
50 Ω | 1 MΩ
Prior to A.02.00
Input/Output, I/Q, Q Setup
[:SENSe]:CORRection:IQ:Q:SKEW <seconds>
[:SENSe]:CORRection:IQ:Q:SKEW?
Delay the data for the Q channel by 10 ns.
CORR:IQ:Q:SKEW 10 ns
0
Yes
This is unaffected by a Preset but is set to the default value on a "Restore
Input/Output Defaults" or "Restore System Defaults->All"
0 s to 100 ns
Prior to A.02.00
See"I/Q Probe Setup" on page
Input/Output, I/Q, Q Setup
No
[<Q port probe id>]
This is reporting the type of probe sensed on the Q port. There is no parameter
for overriding what is sensed.
Prior to A.02.00
424.