Agilent Technologies X Series Manual page 615

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Preset
State Saved
Initial S/W Revision
Screen
This controls the display of the annunciation and annotation around the graticule, including any
annotation on lines (such as the display line, the threshold line, etc.) and the y-axis annotation. This does
NOT include marker annotation (or the N dB result). When off, the graticule expands to fill the entire
graticule area, leaving only the 1.5% gap above the graticule as described in the Trace/Detector chapter.
Key Path
Remote Command
Example
Dependencies
Preset
State Saved
Initial S/W Revision
Trace
Turns on and off the labels on the traces, showing their detector (or their math mode) as described in the
Trace/Detector section.
If trace math is being performed with a trace, then the trace math annotation will replace the detector
annotation.
Key Path
Remote Command
Example
Preset
State Saved
Initial S/W Revision
On
This should remain Off through a Preset when
is set to Off.
Annotation
Saved in instrument state.
Prior to A.02.00
View/Display, Display, Annotation
:DISPlay:ANNotation:SCReen[:STATe] OFF|ON|0|1
:DISPlay:ANNotation:SCReen[:STATe]?
DISP:ANN:SCR OFF
Grayed out and forced to OFF when
set to Off.
On
This should remain Off through a Preset when
is set to Off
Annotation
Saved in instrument state.
Prior to A.02.00
View/Display, Display, Annotation
:DISPlay:ANNotation:TRACe[:STATe] ON|OFF|1|0
:DISPlay:ANNotation:TRACe[:STATe]?
DISP:ANN:TRAC OFF
Off
Saved in instrument state.
Prior to A.02.00
View/Display
System Display Settings
,
System Display Settings
System Display Settings
,
is
Annotation
,
615