System Functions
System
Notes
Initial S/W Revision
Modified at S/W Revision
Query the Operating Temperature Extremes
Returns the low operating temperature extreme value. The value survives a power-cycle and is the temperature
extreme encountered since the value was reset by the factory or service center.
Mode
Remote Command
Example
Notes
State Saved
Initial S/W Revision
Mode
Remote Command
Example
Notes
State Saved
Initial S/W Revision
Query the Elapsed Time since 1st power on
Returns the elapsed on-time in minutes since 1
Remote Command
Example
Notes
206
Query Only
The return value is a comma separated list of the individual counts for each
mechanical relay.
The position of the relays in the list is:
"<Cal Signal>,<AC/DC>,<2dB #1 Atten>,<2dB #2 Atten>,<6dB
Atten>,<10dB Atten>,<20dB Atten>,<30dB Atten>,<Fixed Atten>,<Low
Noise Path Switch>,<Presel Bypass>"
Items in the list not pertaining to your particular hardware configuration will
return as –999 for those items.
Prior to A.02.00
A.03.00, A.04.00
All
:SYSTem:TEMPerature:LEXTreme?
:SYST:TEMP:LEXT?
Value is in degrees Celsius at which the lowest operating temperature has been
recorded since 1st power-up.
No
Prior to A.02.00
All
:SYSTem:TEMPerature:HEXTreme?
:SYST:TEMP:HEXT?
Value is in degrees Celsius at which the highest operating temperature has
been recorded since 1st power-up.
No
Prior to A.02.00
st
power-on.
:SYSTem:PON:ETIMe?
:SYST:PON:ETIM?
Query Only