Notes
Initial S/W Revision
All but RF
Immediately executes an alignment of all subsystems except the RF subsystem. The instrument will stop
any measurement currently underway, perform the alignment, and then restart the measurement from the
beginning (similar to pressing the
are not impacted by an interfering user input signal.
This operation might be chosen instead of
large change in input impedance, such as a temporary open circuit at the analyzer input.
The query form of the remote commands (:CALibration:NRF?) will invoke the alignment and return a
success or failure value.
Successful completion of
Condition, and clear bit 14 in the Status Questionable Calibration register. If "Align Now, All required"
was in effect prior to executing the All but RF, the Error Condition message "Align Now, RF required" is
generated and bit 12 in the Status Questionable Calibration register is set. It will also begin the elapsed
time counter for Last Align Now, All Time, and capture the Last Align Now, All Temperature.
,
can be interrupted by pressing the Cancel (ESC) front-panel key or remotely with
Align Now
All but RF
Device Clear followed by the:ABORt SCPI command. When this occurs the Error Condition message
"Align Now, All required" is generated, and bit 14 is set in the Status Questionable Condition register.
This is because new alignment data may be used for an individual subsystem, but not a full new set of
data for all subsystems.
In models with the RF Preselector, such as the N9038A, the "All but RF" alignment will execute an
alignment of all subsystems except the RF subsystem of the Spectrum Analyzer, as well as the system
gain of the RF Preselector.
Key Path
Mode
Remote Command
Example
*CAL? returns 0 if successful
*CAL? returns 1 if failed
:CALibration[:ALL]? is the same as *CAL?
See additional remarks described with :CALibration[:ALL]?
Everything about :CALibration[:ALL]? is synonymous with *CAL? including
all conditions, status register bits, and couplings
Prior to A.02.00
key). This can be used to align portions of the instrument that
Restart
if you do not want the device under test to experience a
All
Align Now
,
All but RF
will clear the "Align Now, All required" Error
System, Alignments, Align Now
All
:CALibration:NRF
:CALibration:NRF?
:CAL:NRF
System Functions
System
169