Example
Notes
Preset
State Saved
Min
Max
Readback Text
Initial S/W Revision
Q Ext Gain
This function affects only the Q channel input and only when the Input Path is not I+jQ. It is not
available unless the Baseband I/Q option (BBA) is installed.
Key Path
Remote Command
Example
Notes
Preset
State Saved
Min
Max
Readback Text
Initial S/W Revision
Set the I Ext Gain to 10 dB
CORR:IQ:I:GAIN 10
Set the I Ext Gain to –10 dB (that is, a loss of 10 dB.)
CORR:IQ:I:GAIN –10
Not available unless option BBA is installed
0 dB
This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output
Defaults" or "Restore System Defaults->All"
Saved in instrument state.
–100 dB
100 dB
I Gain, <I Ext Gain> dB
Prior to A.02.00
Input/Output, External Gain
[:SENSe]:CORRection:IQ:Q:GAIN <rel_ampl>
[:SENSe]:CORRection:IQ:Q:GAIN?
Set the Q Ext Gain to 10 dB
CORR:IQ:Q:GAIN 10
Set the Q Ext Gain to –10 dB (that is, a loss of 10 dB.)
CORR:IQ:Q:GAIN –10
Not available unless option BBA is installed.
0 dB
This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output
Defaults" or "Restore System Defaults->All"
Saved in instrument state.
–100 dB
100 dB
Q Gain, <I Ext Gain> dB
Prior to A.02.00
Input/Output
435