Input/Output
Key Path
Example
Readback
Initial S/W Revision
Gate
Selects the gate signal to be output to the Trig 1 Out or Trig 2 Out connector. The gate signal has been
delayed and its length determined by delay and length settings. When the polarity is positive, a high on
the Trig 1 Out or Trig 2 Out represents the time the gate is configured to pass the signal.
Key Path
Example
Readback
Initial S/W Revision
Odd/Even Trace Point
Selects either the odd or even trace points as the signal to be output to the Trig 1 Out or Trig 2 Out
connector when performing swept spectrum analysis. When the polarity is positive, this output goes high
during the time the analyzer is sweeping past the first point (Point 0) and every other following trace
point. The opposite is true if the polarity is negative.
Key Path
Example
Readback
Initial S/W Revision
Analog Out
This menu lets you control which signal is fed to the "Analog Out" connector on the analyzer rear panel.
See
"More Information" on page 461
Key Path
Remote Command
Example
Preset
Preset
460
Input/Output, Output Config, Trig 1/2 Output
TRIG1:OUTP GTR
Gate Trigger
Prior to A.02.00
Input/Output, Output Config, Trig 1/2 Output
TRIG1:OUTP GATE
Gate
Prior to A.02.00
Input/Output, Output Config, Trig 1/2 Output
TRIG1:OUTP OEV
Odd/Even
Prior to A.02.00
Input/Output, Output Config
:OUTPut:ANALog OFF|SVIDeo|LOGVideo|LINVideo|DAUDio
:OUTPut:ANALog?
OUTP:ANAL SVIDeo ! causes the analog output type to be Screen Video
OFF
This is unaffected by Preset but is set to DAUDio on a "Restore Input/Output
Defaults" or "Restore System Defaults->All