Complex Spectrum Measurement
Meas Setup
Filter Type
This parameter allows you to select the type of IF filter (post ADC, digital filter) that is used. This is an
advanced control that normally does not need to be changed.
Key Path
Mode
Remote Command
Example
Dependencies
Couplings
Preset
State Saved
Range
Backwards Compatibility SCPI
Initial S/W Revision
Remote Command
Digital IF BW
Allows you to select the type of IF filter (post ADC, digital filter) that is used. This is an advanced
control that normally does not need to be changed.
The Digital IF bandwidth is determined from the Span. The Digital IF BW and Span relation is
determined by the following formula.
Digital IF BW = Span x 1.5
This equation holds except for the case when only the value is clipped to the max value. Furthermore,
248
Meas Setup, Advanced
BASIC
[:SENSe]:SPECtrum:DIF:FILTer:TYPE
GAUSsian|FLATtop|SNYQuist|RSNYquist|RCOSine|RRCosine
[:SENSe]:SPECtrum:DIF:FILTer:TYPE?
SPEC:DIF:FILT:TYPE GAUS
SPEC:DIF:FILT:TYPE?
Gaussian and Flattop are available in all DIF configurations. For the other
filter types, the filters are only available with PXA or when Option B40 is
installed.
Based on the DIF board and the IF Path Selection parameter which is
described in the table above.
FLAT
Saved in instrument state.
Gaussian | Flattop
(With Digital IF and/or Option B40 or B1X)
Gaussian|Flattop|Short nyquist|Root Short Nquist|Raised Cosine|Root
RaisedCosine
[:SENSe]:SPECtrum:BANDwidth:PFFT:TYPE
[:SENSe]:SPECtrum:BWIDth:PFFT:TYPE
[:SENSe]:SPECtrum:BANDwidth|BWIDth:IF:SHAPe
Prior to A.02.00
[:SENSe]:SPECtrum:WBIF:FILTer[:TYPE] GAUSsian|
NONE|NYQuist|RNYQuist|RCOSine|RRCosine
[:SENSe]:SPECtrum:WBIF:FILTer[:TYPE]?