AMPTD Y Scale
Min
Max
Initial S/W Revision
Q Range
Accesses the Q Range menu.
Key Path
Readback Text
Initial S/W Revision
Q Same as I
Many, but not all, usages require the I and Q channels to have an identical setup. To simplify channel
setup, the Q Same as I will cause the Q channel range to be mirrored from the I channel. That way you
only need to set up one channel (the I channel). The I channel values are copied to the Q channel, so at
the time Q Same as I is Off, the I and Q channel setups will be identical.
Key Path
Remote Command
Example
Couplings
Preset
State Saved
Range
Readback Text
Initial S/W Revision
Q Range Value
This is the internal gain range for the Q channel. See
applies to Input Path Q Only and Ind I/Q. For input I+jQ the I Range determines both I and Q channel
range settings.
Key Path
364
–20 dBm
10 dBm
Prior to A.02.00
AMPTD Y Scale, Range
Q Same as I | 1 V Peak | 0.5 V Peak | 0.25 V Peak | 0.125 V Peak
When Q Same as I is On, the readback is "Q Same as I", otherwise it is the Q
Range value.
Prior to A.02.00
AMPTD Y Scale, Range, Q Range
[:SENSe]:VOLTage|POWer:IQ:MIRRored OFF|ON|0|1
[:SENSe]:VOLTage|POWer:IQ:MIRRored?
Turn off the mirroring of I Range to Q Range.
VOLT:IQ:MIRR OFF
POW:IQ:MIRR OFF
When On, the I Range value is mirrored (copied) to the Q Range.
On
Saved in instrument state.
On | Off
"Q Same as I" when On, otherwise none.
Prior to A.02.00
"I/Q Gain Ranges" on page
AMPTD Y Scale, Range
366. The Q Range only