A sample of the Gate View screen in the Swept SA measurement is shown in the following graphic:
550
Sweep/Control, Gate
[:SENSe]:SWEep:EGATe:VIEW ON|OFF|1|0
[:SENSe]:SWEep:EGATe:VIEW?
SWE:EGAT:VIEW ON turns on the gate view.
In the Swept SA measurement:
In Gate View, the regular Sweep Time key is grayed out. When pressed, the
grayed out key puts up the informational message "Use Gate View Sweep
Time in the Gate menu."
In the other measurements:
When you turn Gate View on, the lower window takes on the current state of
the instrument. Upon leaving Gate View, the instrument takes on the state of
the lower window.
When you turn Gate View on, the upper window Sweep Time is set to the gate
view sweep time.
These couplings apply to the Swept SA measurement:
When Gate View is turned on, the instrument is set to Zero Span.
Gate View automatically turns off whenever a Span other than Zero is
selected.
Gate View automatically turns off if you press the Last Span key while in
Gate View, and the instrument returns to the Span it was in before entering
Gate View (even if that is Zero Span).
When Gate View is turned on, the sweep time used is the gate view sweep
time. This is set according to the rules in
page
552.
When Gate View is turned off, Sweep Time is set to the normal Swept SA
measurement sweep time.
If Gate View is on and Gate is off, then turning on Gate turns off Gate
View.
OFF
Saved in instrument state
On|Off
Prior to A.02.00
"Gate View Sweep Time" on