System Functions
System
Remote Command
Example
Preset
Preset
State Saved
Range
Initial S/W Revision
SCPI Socket Control Port (Remote Command Only)
Returns the TCP/IP port number of the control socket associated with the SCPI socket session. This
query enables you to obtain the unique port number to open when a device clear is to be sent to the
instrument. Every time a connection is made to the SCPI socket, the instrument creates a peer control
socket. The port number for this socket is random. The user must use this command to obtain the port
number of the control socket. To force a device clear on this socket, open the port and send the string
"DCL " to the instrument.
If this SCPI command is sent to a non SCPI Socket interface, then 0 is returned.
Mode
Remote Command
Example
Preset
State Saved
Range
Initial S/W Revision
SICL Server
Turns the SICL server capability On or Off, enabling you to limit SCPI access over LAN through the
SICL server. (SICL IEEE 488.2 protocol.)
Parameter
Maximum Connections
Instrument Name
Instrument Logical Unit
190
:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle OFF|ON|0|1
:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle?
:SYST:COMM:LAN:SCPI:SOCK:ENAB OFF
ON
This is unaffected by a Preset but is set to ON with a "Restore System
Defaults->Misc"
No
On | Off
Prior to A.02.00
All
:SYSTem:COMMunicate:LAN:SCPI:SOCKet:CONTrol?
:SYST:COMM:LAN:SCPI:SOCK:CONT?
This is unaffected by Preset or "Restore System Defaults->Misc".
No
0 to 65534
Prior to A.02.00
Description
The maximum number of connections that can be accessed
simultaneously
The name (same as the remote SICL address) of your
analyzer
The unique integer assigned to your analyzer when using
SICL LAN
Setting
5
inst0
8