Delete All Corrections
Erases all correction values for all 4 Amplitude Correction sets.
When this key is pressed a prompt is placed on the screen that says "Please press Enter or OK key to
delete all corrections. Press ESC or Cancel to close this dialog." The deletion is only performed if you
press OK or Enter.
Key Path
Remote Command
Example
Initial S/W Revision
Remote Correction Data Set Commands
Set (Replace) Data (Remote Command Only)
The command takes an ASCII series of alternating frequency and amplitude points, each value separated
by commas.
The values sent in the command will totally replace all existing correction points in the specified set.
An Ampcor array can contain 2000 points maximum.
Remote Command
Example
Preset
State Saved
Min
Max
Initial S/W Revision
Merge Correction Data (Remote Command Only)
The command takes an ASCII series of alternating frequency and amplitude points, each value separated
by commas. The difference between this command and Set Data is that this merges new correction points
into an existing set.
Input/Output, Corrections
[:SENSe]:CORRection:CSET:ALL:DELete
CORR:CSET:ALL:DEL
A.02.00
[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DATA <freq>,
<ampl>, . . .
[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DATA?
CORR:CSET1:DATA 10000000,–1.0,20000000,1.0
This defines two correction points at (10 MHz, –1.0 dB) and (20 MHz, 1.0
dB) for correction set 1.
Empty after Restore Input/Output Defaults. Survives a shutdown or restart of
analyzer application (including a power cycle).
Saved in instrument state.
Freq: 0 Hz
Amptd: –1000 dBm
Freq: 1 THz
Amptd: +1000 dBm
A.02.00