Sweep/Control
Sweep Time Rules
Allows the choice of three distinct sets of sweep time rules. These are the rules that are used to set the
sweep time when
Sweep Time
(either manually or automatically chosen) and not when in FFT sweeps.
Sweep Type
See
"More Information" on page
Key Path
Remote Command
Example
Dependencies
Couplings
Preset
State Saved
Backwards Compatibility SCPI
Initial S/W Revision
538
is in
Auto mode.
Note that these rules only apply when in the Swept
539.
Sweep/Control, Sweep Setup
[:SENSe]:SWEep:TIME:AUTO:RULes
NORMal|ACCuracy|SRESponse
[:SENSe]:SWEep:TIME:AUTO:RULes?
SWE:TIME:AUTO:RUL ACC
In Zero Span, this key is irrelevant and cannot be accessed (because the whole
Sweep Setup menu is grayed out in Zero Span), however its settings can be
changed remotely with no error indication.
Grayed out in FFT sweeps. Pressing the key while the instrument is in FFT
sweep generates an advisory message. The SCPI is acted upon if sent, but has
no effect other than to change the readout on the key, as long as the analyzer is
in an FFT sweep.
Set to Auto on Auto Couple
AUTO
Saved in instrument state
The old Auto Sweep Time command was the same
[:SENSe]:SWEep:TIME:AUTO:RULes NORMal|ACCuracy
so it still works although it now has a third parameter (SRESponse).
The old Sweep Coupling command was
[:SENSe]:SWEep:TIME:AUTO:MODE SRESponse|SANalyzer and it is
aliased as follows:
:SWEep:TIME:AUTO:MODE SRESponse is aliased to
:SWEep:TIME:AUTO:RULes SRESponse, and
:SWEep:TIME:AUTO:MODE SANalyzer is aliased to
:SWEep:TIME:AUTO:RULes NORMal
The query :SWEep:TIME:AUTO:MODE? is aliased to
:SWEep:TIME:RULes?
So it will fail to match for SANalyzer
Prior to A.02.00