Mode
Example
Readback Text
Initial S/W Revision
Last State
Sets
to Last. When the analyzer is powered on, it will put all modes in the last state they were
Power On
in prior to when the analyzer was put into Power Standby and it will wake up in the mode it was last in
prior to powering off the instrument. The saving of the active mode prior to shutdown happens behind
the scenes when a controlled shutdown is requested by using the front panel power
using the remote command SYSTem:PDOWn. The non-active modes are saved as they are deactivated
and recalled by Power On Last State.
An instrument could never power up for the first time in Last.
NOTE
If line power to the analyzer is interrupted, for example by pulling the line cord
plug or by switching off power to a test rack, Power On Last State will not work
properly.
Key Path
Mode
Example
Notes
Readback Text
Initial S/W Revision
Power On Application
Accesses a menu that lists the available Modes and lets you select which Mode is to be the power-on
application.
This application is used for Power On Type "Mode and Input/Output Defaults" and Restore System
Defaults All.
Key Path
Mode
All
SYST:PON:TYPE USER
User Preset
Prior to A.02.00
System, Power On
All
SYST:PON:TYPE LAST
Power on Last State only works if the user has done a controlled shutdown
prior to powering on in Last. If a controlled shutdown is not done when in
Power On Last State, the instrument will power up in the last active mode, but
it may not power up in the active mode's last state. If an invalid mode state is
detected, a Mode Preset will occur. To control the shutdown under remote
control use the :SYSTem:PDOWn command.
Last State
Prior to A.02.00
System, Power On
All
System Functions
System
key or by
Standby
153