Agilent Technologies X Series Manual page 558

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Sweep/Control
Gate Period
begins
Gate Delay
Gate
Gate
Signal
Gate signal
goes true
When Gate Holdoff is in Auto, the wait time calculated by the analyzer is used. When Gate Time is in
Manual, the user may adjust the wait time, usually decreasing it in order to achieve greater speed, but at
the risk of decreasing accuracy.
When the
Method
key is set to
In measurements that do not support Auto, the value shown when Auto is selected is "---"and the
manually set holdoff is returned to a query.
Key Path
Remote Command
Example
558
Gate Period
ends
Gate Holdoff
Gate Length
Interval
Gate signal
goes true but
has no effect as
it is during the
Gate Holdoff
interval
Video
or
FFT
, the
Gate Holdoff
Sweep/Control, Gate
[:SENSe]:SWEep:EGATe:HOLDoff <time>
[:SENSe]:SWEep:EGATe:HOLDoff?
[:SENSe]:SWEep:EGATe:HOLDoff:AUTO OFF|ON|0|1
[:SENSe]:SWEep:EGATe:HOLDoff:AUTO?
SWE:EGAT:HOLD 0.0002
SWE:EGAT:HOLD?
SWE:EGAT:HOLD:AUTO ON
SWE:EGAT:HOLD:AUTO?
Gate Period
begins
Gate Delay
Gate signal
goes true
vsd 39-2
function has no effect.