Range
Initial S/W Revision

Auto/Holdoff

Opens up a menu that lets you adjust Auto Trigger and Trigger Holdoff parameters
Key Path
Readback line
Initial S/W Revision
Auto Trig
Sets the time that the analyzer will wait for the trigger conditions to be met. If they are not met after that
much time, then the analyzer is triggered anyway.
Key Path
Remote Command
Example
Notes
Preset
State Saved
Min
Max
10 Hz to Maximum
Prior to A.02.00
Trigger
Displays a summary of the Auto Trig and Holdoff settings, in square brackets
First line: Auto Off or Auto On
Second Line: "Hldf" followed by:
If Holdoff is Off, readback Off
If Holdoff On and Type = Normal, readback value
If Holdoff On and Type = Above, readback value followed by AL
If Holdoff On and Type = Below, readback value followed by BL
If Holdoff Type selection is not supported by the current measurement,
Holdoff Type is always Normal
A.02.00
Trigger, Auto/Holdoff
:TRIGger[:SEQuence]:ATRigger <time>
:TRIGger[:SEQuence]:ATRigger?
:TRIGger[:SEQuence]:ATRigger:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:ATRigger:STATe?
TRIG:ATR:STAT ON
TRIG:ATR 100 ms
The "time that the analyzer will wait" starts when the analyzer is ready for a
trigger, which may be hundreds of ms after the data acquisition for a sweep is
done. The "time" ends when the trigger condition is satisfied, not when the
delay ends.
Off, 100 ms
Saved in instrument state
1 ms
100 s
Trigger
609