Agilent Technologies X Series Manual page 610

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Trigger
Default Unit
Initial S/W Revision
Trig Holdoff
Sets the holdoff time between triggers. When the trigger condition is satisfied, the trigger occurs, the
delay begins, and the holdoff time begins. New trigger conditions will be ignored until the holdoff time
expires. For a free-running trigger, the holdoff value is the minimum time between triggers.
Key Path
Remote Command
Example
Preset
State Saved
Min
Max
Default Unit
Initial S/W Revision
Holdoff Type
Lets you set the Trigger Holdoff Type.
Holdoff Type is not supported by all measurements. If the current measurement
NOTE
does not support it, this key will be blank and the Holdoff Type will be Normal. If
the Holdoff Type SCPI is sent while in such a measurement, the SCPI will be
accepted and the setting remembered, but it will have no effect until a
measurement is in force that supports Holdoff Type.
Trigger Holdoff Type functionality:
• NORMal
This is the "oscilloscope" type of trigger holdoff, and is the setting when the Holdoff Type key does
not appear. In this type of holdoff, no new trigger will be accepted until the holdoff interval has
expired after the previous trigger.
610
s
Prior to A.02.00
Trigger, Auto/Holdoff
:TRIGger[:SEQuence]:HOLDoff <time>
:TRIGger[:SEQuence]:HOLDoff?
:TRIGger[:SEQuence]:HOLDoff:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:HOLDoff:STATe?
TRIG:HOLD:STAT ON
TRIG:HOLD 100 ms
Off, 100 ms
Saved in instrument state
0 s
0.5 s
s
Prior to A.02.00