Operation Status Event Register (:Operegister[:Event]) - Agilent Technologies InfiniiVision 2000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 2000 X-Series:
Table of Contents

Advertisement

35
Status Reporting

Operation Status Event Register (:OPERegister[:EVENt])

The Operation Status Event Register register hosts these bits:
Name
RUN bit
WAIT TRIG bit
MTE bit
OVLR bit
If any of these bits are set, the OPER bit (bit 7) of the Status Byte
Register is set. The Operation Status Event Register is read and cleared
with the :OPERegister[:EVENt]? query. The register output is enabled or
disabled using the mask value supplied with the OPEE command.
772
Location
Description
bit 3
Is set whenever the instrument goes from a stop
state to a single or running state.
bit 5
Is set by the Trigger Armed Event Register and
indicates that the trigger is armed.
bit 9
Comes from the Mask Test Event Registers.
bit 11
Is set whenever a 50 input overload occurs.
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide

Advertisement

Table of Contents
loading

Table of Contents