CALibration Subsystem

CALibration[1]:AUTO [ONCE|ON|OFF|0|1]

This command zeroes the power sensor when enabled.
When ONCE is enabled, the zeroing will perform only once for both internal
and external zeroing type.
When 1|ON is enabled, the zero is updated if the sensor's temperature
changes by
internal zeroing. The zero is maintained by on- the- fly zero measurements.
The 0|OFF parameter is only required for the query response and is ignored in
the command.
This command performs the same function as CALibration:ZERO:AUTO
N O T E
[ONCE|ON|OFF|0|1].
Syntax
Example
CAL1:AUTO ONCE
112
o
±
5
C or the time since last zeroing is greater than 24 hours during
CAL
1
:AUTO
Space
0|OFF
1|ON
ONCE
?
This command causes the power sensor to
perform a zeroing.
U2000 Series Programming Guide