Parameters
Item
Description/Default
source
Available trigger sources:
BUS: the trigger source is a *TRG common command or the
•
TRIGGER:IMMediate SCPI command.
EXTernal: the trigger source is the trigger input in the sensor.
•
HOLD: triggering is suspended. The only way to trigger the power sensor is to
•
use TRIGger:IMMediate.
IMMediate: the trigger system is always true. If INITiate:CONTinuous
•
is ON the power sensor is continually triggering free (free run mode). If an
INITiate:IMMediate command is sent a measurement is triggered then
the power sensor returns to the idle state.
TThe trigger source is set to IMMediate on instrument power-up.
N O T E
The MEASure and CONFigure commands automatically set the trigger source to
IMMediate.
The READ? or MEASure commands should not be used if the trigger source is set to HOLD.
Example
TRIG:SOUR IMM
U2000 Series Programming Guide
This command configures the sensor for
immediate triggering.
TRIGger Subsystem
Range of Values
EXTernal
HOLD
IMMediate
BUS