7
SENSe Subsystem
Parameters
Example
AVER:COUN 400
Reset Condition
On reset, the filter length is set to four.
Query
AVERage:COUNt? [MIN|MAX]
The query returns the current setting of the filter length or the values
associated with MIN and MAX. The format of the response is <NR1>.
164
Item
Description/Default
numeric_value
A numeric value defining the filter length.
DEF: the default value is 4
MIN: 1
MAX: 1024
This command enters a filter length of 400
for power sensor.
U2000 Series Programming Guide
Range of Values
1 to 1024
DEF
MIN
MAX