Agilent Technologies U2000 Series Manual page 193

Usb power sensors
Hide thumbs Also See for U2000 Series:
Table of Contents

Advertisement

7
SENSe Subsystem
Example
CORR:CSET2 'PW1'
Reset Condition
On reset the selected table is not affected.
Query
SENSe[1]:CORRection:CSET2:[SELect]?
The name of the selected table is returned as a quoted string. If no table is
selected an empty string is returned.
Query Example
CORR:CSET2?
Error Messages
• If <string> is not valid, error –224, "Illegal parameter value" occurs.
• If a table called <string> does not exist, error –256, "File name not found"
175
occurs.
When a frequency dependent offset table is selected, the power
sensorpower sensor verifies that the number of offset points defined is
equal to the number of frequency points defined. If this is not the case,
error –226, "Lists not the same length" occurs.
This command enters the name of the
sensor frequency dependent offset table
which is to be used.
This command queries the sensor
frequency dependent offset table currently
used.
U2000 Series Programming Guide