Power Sensor Remote Operation
1
READ? Example
ABORt1
CONFigure1:POWer:AC DEF,2,(@1)
SENS1:CORR:CSET2:SEL "Offset1"
SENS1:CORR:CSET2:STAT ON
SENS1:FREQuency 500KHZ
READ1?
If the measurement frequency does not correspond directly to a frequency in the frequency
N O T E
dependent offset table, the power sensor calculates the offset using linear interpolation.
If you enter a frequency outside the frequency range defined in the frequency dependent
offset table, then the power sensor uses the highest or lowest frequency point in the table
to set the offset.
To find out the value of the offset being used by the power sensor to make a measurement,
use the query command:
[SENSe[1]]:CORRection:GAIN4|FDOFfset[:INPut][MAGNITUDE]?
The response may be an interpolated value.
U2000 Series Programming Guide
23