1
Power Sensor Remote Operation

Saving and Recalling Power Sensor Configurations

To reduce repeated programming, up to ten power sensor configurations can
be stored in the non- volatile memory. The error list, zeroing and calibration
information are not stored.

How to Save and Recall a Configuration

Power sensor configurations are saved and recalled with the following
commands:
*SAV <NRf>
*RCL <NRf>
The range of values for <NRf> in the above commands is 1 to 10.