Status:operation:calibrating[:Summary] - Agilent Technologies U2000 Series Manual

Usb power sensors
Hide thumbs Also See for U2000 Series:
Table of Contents

Advertisement

9
STATus Subsystem

STATus:OPERation:CALibrating[:SUMMary]

The operation status calibrating summary register set contains information on
the calibrating status of the power sensor.
The following bits in these registers are used by the power sensor:
These bits are set at the beginning of zeroing (CALibration:ZERO:AUTO
ONCE) and at the beginning of calibration (CALibration:AUTO ONCE). Also for
the compound command/query CALibration[:ALL]?, this bit is set at the
beginning of the calibration sequence.
These bits are cleared at the end of zeroing or calibration.
Syntax
234
Bit Number
Decimal
Weight
0
-
1
2
2
4
3-15
-
:OPER
STAT
Definition
Not used
CALibrating Status
Not used
Not used
:SUMM
:CAL
U2000 Series Programming Guide