U2000 Series Programming Guide
Block data not allowed
A legal block data element was encountered but not allowed by the power sensor at
this point.
For example SYST:LANG #15FETC?.
Expression data not allowed
A legal expression data was encountered but not allowed by the power sensor at
this point.
For example SYST:LANG (5+2).
Trigger ignored
Indicates that *TRG, or TRIG:IMM was received and recognized by the device but
was ignored because the power sensor was not in the wait for trigger state.
Init ignored
Indicates that a request for a measurement initiation was ignored as the power
sensor was already initiated.
For example, INIT:CONT ON
INIT.
Trigger deadlock
TRIG:SOUR was set to HOLD or BUS and a READ? or MEASure?
was attempted, expecting TRIG:SOUR to be set to IMMediate.
Parameter error;Frequency list must be in ascending order.
Indicates that the frequencies entered using the
MEMory:TABLe:FREQuency command are not in ascending order.
Settings conflict
This message occurs under a variety of conflicting conditions. The following list
gives a few examples of where this error may occur:
If the READ? parameters do not match the current settings.
If you are in fast mode and attempting to switch on for example,
averaging, duty cycle or limits.
Trying to clear a sensor calibration table when none is selected.
Data out of range
A numeric parameter value is outside the valid range for the command.
For example, SENS:FREQ 2KHZ.
Illegal parameter value
A discrete parameter was received which was not a valid choice for the command.
You may have used an invalid parameter choice.
For example, TRIG:SOUR EXT.
SYSTem Subsystem