3
CALCulate Subsystem

CALCulate[1]:LIMit:FCOunt?

This query returns the total number of limit failures for the current
window/measurement.
If the appropriate STATe commands are set to ON, each time a measurement is
initiated on the current window/measurement and the result is outside the
limits, the counter is incremented by one.
If the measured value is equal to a limit, this is a limit pass.
The counter is reset to zero by any of the following commands:
• *RST
• CALCulate[1]:LIMit:CLEar:IMMediate
• CALCulate[1]:LIMit:CLEar:AUTO ON
When CALCulate[1]:LIMit:CLEar:AUTO is set to ON, the counter is set to
zero each time a measurement is:
• measured using MEASure?
• read using READ?
• initiated using:
When CALCulate[1]:LIMit:CLEar:AUTO is set to ONCE, the counter is set to
zero the first time a measurement is:
• measured using MEASure?
• read using READ?
• initiated using:
The maximum number of errors is 2
detected the counter returns to zero.
94
INITiate[:IMMediate] or,
INITiate:CONTinuous ON
INITiate[:IMMediate] or,
INITiate:CONTinuous ON
16
16
–1. If more than 2
U2000 Series Programming Guide
–1 errors are