Interconnections And Calibration - Agilent Technologies E Series Operating And Service Manual

Power sensors
Hide thumbs Also See for E Series:
Table of Contents

Advertisement

Introduction
Getting Started

Interconnections and Calibration

Connect one end of an Agilent 11730 series sensor cable to the Agilent E-
series E9300 power sensor and connect the other end of the cable to the
power meter's channel input. Allow a few seconds for the power meter to
download the power sensor's calibration table.
Caution
The Agilent 9304A Sensor is DC coupled. DC voltages in excess of the maximum value
(5 Vdc) can damage the sensing diode.
Note
Ensure power sensors and cables are attached and removed in an indoor environment.
To carry out a zero and calibration cycle as requested by the power meter
proceed as follows:
On completion the power meter and sensor are ready to connect to the
device under test (DUT). Ensure the attenuator is re- connected to the
Agilent E- series E9300B or E9301B sensors prior to making measurements.
Caution
The Agilent E-series E9300B or E9301B sensors should not be operated without the
attenuator connected at any time other than for calibration. You must ensure the
attenuator is reconnected following calibration.
18
Ensure the Agilent E- series E9300 power sensor is disconnected
from any signal source.
When calibrating Agilent E- series E9300B or E9301B sensors, first
remove the attenuator.
On the power meter, press
During zeroing the wait symbol is displayed.
When the wait period is complete connect the Agilent E- series
power sensor to the power meter's POWER REF output.
Cal
Cal Cal A
Press
(or
displayed during calibration.
Zero
Zero
,
Cal
Cal B
,
/
). The wait symbol is again
Zero A
Zero B
(or
/
).