Using READ?
ABOR1
CONF1 -50,DEF,(@2)
READ1?
Some fine tuning of measurements can be performed using the CONFigure and
READ? commands. For example, in the above program segment some fine
tuning can be performed by setting the filter length to 1024 and the trigger
delay off.
1 ABOR1
2 CONF1 -50,DEF,(@1)
3 SENS1:AVER:COUN 1024
4 TRIG1:DEL:AUTO OFF
5 READ1?
Using INITiate and FETCh?
ABOR1
CONF1 -50,DEF,(@1)
INIT1
FETC1? -50,DEF,(@1)
Some fine tuning of measurements can be carried out using the CONFigure
command and INITiate and FETCh? commands. For example, in the above
program segment some fine tuning can be carried out by setting the filter
length to 1024 and the trigger delay off.
1 ABOR1
2 CONF1 -50,DEF,(@1)
3 SENS1:AVER:COUN 1024
U2000 Series Programming Guide
Power Sensor Remote Operation
Aborts current measurement
Configures the current measurement to use
an expected power of –50 dBm and the
current resolution setting
Takes the current measurement
Aborts current measurement
Configures the current measurement to use
an expected power of –50 dBm and the
current resolution setting
Causes a measurement
Retrieves the current measurement