SENSe Subsystem
SENSe[1]:CORRection:CSET[2]:STATe <boolean>
This command is to enable and disable the use of the currently active sensor
frequency dependent offset table (CSET2). When a table has been selected and
enabled, the sensor frequency dependent offsets stored in it can be used by
specifying the required frequency using the SENSe[1]:FREQuency command.
When the CSET2 command is set to ON, the frequency dependent offset is taken
from the sensor frequency dependent offset table and is used.
Syntax
SENS
1
Example
CORR:CSET2:STAT 1
Reset Condition
On reset, the sensor frequency dependent offset table are not affected.
Query
SENSe[1]:CORRection:CSET[2]:STATe?
The query returns a 1 or 0 into the output buffer indicating whether a table is