2
MEASurement Commands

MEASure[1] Commands

The MEASure? commands configure the power sensor to perform a power
measurement with the given measurement function, range and resolution then
makes the measurement. The format of the result is set by
FORM[:READ][:DATA]. Refer to
for further information.
MEASure? is a compound command which is equivalent to:
• The MEASure? query is equivalent to:
80
ABORt
CONFigure
READ?
or
ABORt1
CONFigure1
READ1?
Chapter
5, "FORMat Subsystem," on page 119
U2000 Series Programming Guide