Setting Limits (Check Header - same as previous)
The power sensor can be configured to verify the current measurement in any
measurement line against predefined upper and/or lower limit values. The
range of values that can be set for the upper and lower limits and the default
values depends on the measurement units in the currently measurement line -
see
Table 1-2 Range of Values for Measurement Limits
Checking for Limit Failures
There is a way to check for limit failures:
• Use the CALCulate[1]:LIMit:FAIL? and CALCulate[1]:LIMit:FCOunt?
Using CALCulate
Using CALCulate to check the measurement limit failures in
would return the following results:
CALCulate[1]:LIMit:FAIL?
CALCulate[1]:LIMit:FCOunt?
U2000 Series Programming Guide
Table 1-
2.
Window
Units
Maximum
dB
+200 dB
dBm
+230 dBm
%
999.9 X%
W
100.000 XW
commands for measurement limits
Power Sensor Remote Operation
Minimum
Maximum
–
60 dB
180 dB
–
90 dBm
150 dBm
100.0 a%
100.0 M%
1.000 aW
1.000 MW
Returns 1 if there has been 1 or
more limit failures or 0 if there
have been no limit failures. In this
case 1 is returned.
Returns the total number of limit
failures, in this case 2.
1
Default
Minimum
–
120 dB
–
90 dBm
100.0 p%
1.000 pW
Figure 1- 4
29