INITiate[1][:IMMediate]:SEQuence[1]

This command sets the power sensor in the wait for trigger state. When a
trigger is received, the measurement is taken and the result placed in the
power sensor memory. If TRIGger:SOURce is set to IMMediate the
measurement begins as soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer.
Refer to
This command performs the same function as INITiate[1]:[IMMediate].
N O T E
Syntax
Example
INIT:IMM:SEQ1
Error Messages
If the power sensor is not in the "idle" state or INITiate:CONTinuous is ON,
error –213, "INIT ignored" occurs.
U2000 Series Programming Guide
"FETCh[1]?"
on page 74 for further information.