SERVice:SENSor[1]:RADC?
This query returns a new raw uncorrected measurement in volts, as a 32 bit
signed integer.
Syntax
Example
SERV:SENS1:RADC?
Returned raw measurement is in float type unit.
N O T E
Error Messages
• If INIT:CONT is set to ON, error –221 "Settings Conflict" occurs.
U2000 Series Programming Guide
SERV
:SENS
:RADC
1
This query returns a new raw
uncorrected measurement for the sensor.
SERVice Subsystem
?
8