STATus:OPERation

The operation status register set contains conditions which are a part of the
operation of the power sensor as a whole.
The following bits in these registers are used by the power sensor:
Syntax
U2000 Series Programming Guide
Bit Number
Decimal
Weight
0
1
1 - 3
-
4
16
5
32
6 - 9
-
10
1024
11
2048
12
4096
13 to 15
-
:OPER
STAT
Definition
CALibrating Summary
Not used
MEASuring Summary
Waiting for TRIGger Summary
Not used
SENSe Summary
Lower Limit Fail Summary
Upper Limit Fail Summary
Not used (bit 15 always 0)