Power Sensor Remote Operation
an error occurs.
Using READ?
ABOR1
CONF1 DEF,DEF,(@1)
READ1?
Using INITiate and FETCh?
ABOR1
CONF1 DEF,DEF,(@1)
INIT1
FETC1? DEF,DEF,(@1)
Example 3 - Specifying the Expected Power Parameter
The previous example details the three optional parameters which can be used
with the CONFigure and READ? commands. The first optional parameter is
used to enter an expected power value. The value entered determines which of
the power sensor's two ranges is used for the measurement. If the current
setting of the power sensor's range is no longer valid for the new
measurement, specifying the expected power value decreases the time taken
to obtain a result.
The following example uses the expected value parameter to specify a value of
–50 dBm. This selects the power sensor's lower range (refer to
page 26 for details of the range breaks). The resolution parameter is defaulted,
leaving it at its current setting.
12
Aborts measurement
Configures the current window to make a
measurement using the current expected
power and resolution settings
Takes the current window's measurement
Aborts measurement
Configures the current window to make a
measurement using the current expected
power and resolution settings
Causes a measurement
Retrieves the current window's
measurement
U2000 Series Programming Guide