CALibration Subsystem

CALibration Subsystem

The CALibration command subsystem is used to zero the power sensor.
Zeroing of the power sensor is recommended:
• When a 5
• When you change the power sensor
• Every 24 hours
• Prior to measuring low level signals. For example, 10 dB above the lowest
The following CALibration commands are overlapped commands:
• CAL:ALL
• CAL:AUTO
• CAL:ZERO:AUTO
An overlapped command allows the instrument to continue parsing and
executing subsequent commands while it is still executing.
Keyword
CALibration
[:ALL]
[:ALL]?
:AUTO
:ZERO
108
o
C change in temperature occurs
specified power for your sensor.
Parameter Form
<boolean>|ONCE
:AUTO
<boolean>|ONCE
:TYPE
<EXTernal|INTernal>
Notes
[event; no query]
[event; query]
U2000 Series Programming Guide
Page
page 109
page 110
page 112
page 114
page 116