Agilent Technologies U2000 Series Manual page 211

Usb power sensors
Hide thumbs Also See for U2000 Series:
Table of Contents

Advertisement

7
SENSe Subsystem
Parameters
Example
FREQ 500kHz
Reset Condition
On reset, the frequency is set to 50 MHz (DEF).
193
Item
Description/Default
numeric_value
A numeric value for the frequency:
The default units are Hz.
1
The following measurement units can be used:
• Hz
3
• kHz (10
)
6
• MHz (10
)
9
• GHz (10
)
DEF: the default value is 50 MHz
MIN: 1 kHz
MAX: 1000.0 GHz
This command enters a sesnor frequency
of 500 kHz.
Range of Values
1 kHz to 1000.0 Ghz
DEF
MIN
MAX
U2000 Series Programming Guide
1

Hide quick links:

Advertisement

Table of Contents
loading