Agilent Technologies U2000 Series Manual page 97

Usb power sensors
Hide thumbs Also See for U2000 Series:
Table of Contents

Advertisement

Example
READ1:POW:AC?
Error Messages
• INITiate:CONTinuous must be set to OFF, otherwise error –213, "INIT
• If TRIGger:SOURce is set to BUS or HOLD, error –214, "Trigger deadlock"
• If the expected value and resolution parameters are not the same as the
U2000 Series Programming Guide
Item
Description/Default
resolution
A numeric value for the resolution. If it is
unspecified the current resolution setting is
used. If a value is entered it should
correspond to the current resolution setting
otherwise an error occurs.
source list
The measurement which the command is
implemented on.
If unspecified the current window setup is
used.
1
The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the
command sub-systems. The parameters must be entered in the specified order. If parameters are
omitted, they default from the right. The parameter DEFault is used as a place holder. Specifying
DEF leaves the parameter value unchanged.
2
When the measurement result is linear this parameter represents the number of significant
digits. When the measurement result is logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001
respectively.
ignored" occurs.
occurs.
current expected value and resolution settings on the current window,
error –221, "Settings conflict" occurs.
MEASurement Commands
This command queries the current
window/measurement.
2
Range of Values
2
1 to 4
1.0, 0.1, 0.01, 0.001
1
DEF
(@1)
79