Power Sensor Remote Operation

Filter Length

You specify the filter length using the following command:
[SENSe[1]]:AVERage:COUNt <numeric_value>
The range of values for the filter length is 1 to 2048. Specifying this command
disables automatic filter length selection. Increasing the value of the filter
length reduces measurement noise. However, the time to take the
measurement is increased.

Range

With an U2000 Series USB Power Sensor the range can be set either
automatically or manually. Use autoranging when you are not sure of the
power level you will be measuring.
Setting the Range
To set the range manually use the following command:
[SENSe[1]]:POWer:AC:RANGe <numeric_value>
If the <numeric_value> is set to:
• 0, the sensor's lower range is selected.
• 1, the sensor's upper range is selected.
For detailed on the range limits of U2000 Series USB power sensors, refer to
the U2000 Series USB Power Sensors Operating and Service Guide.
For further information on this command refer to
To enable autoranging use the following command:
[SENSe[1]]:POWer:AC:RANGe:AUTO ON
Use autoranging when you are not sure of the power level you will be
measuring.