Operation Status
The Operation Status group monitors conditions in the power sensor's
measurement process.
The Operation status model is shown in the pullout at the end of this chapter.
The following bits in these registers are used by the power sensor:

Table 1-7 Bit Definitions - Operation Status

U2000 Series Programming Guide
Bit
Meaning
Number
8
CALibration
Summary
9
Power On Self
Test
Bit Number
Decimal
Weight
0
1
1 - 3
-
4
16
5
32
6 - 9
-
10
1024
11
2048
Power Sensor Remote Operation
EVENts Causing Bit Changes
This is a summary bit for the Questionable CALibration
Register.
SET:
These may be caused by
CALibration[1]:ZERO:AUTO ONCE or
CALibration[1]:AUTO ONCE or
CALibration[1][:ALL] or
CALibration[1][:ALL]?.
Error –231, "Data questionable; ZERO ERROR"
Error –231, "Data questionable; CAL ERROR"
CLEARED: When any of the commands listed above
succeed and no errors are placed on the error queue.
SET: This bit is set when the power on self test fails.
CLEARED: When the power on self test passes.
Definition
CALibrating Summary
Not used
MEASuring Summary
Waiting for TRIGger Summary
Not used
SENSe Summary
Lower Limit Fail Summary
1
45