11
TRIGger Subsystem
Example
TRIG:SEQ1:COUN 10
Reset Condition
On reset, the value is set to 1.
Query
TRIGger:SEQuence[1]:COUNt?
The query returns the current setting of trigger events for the sensor.
Query Example
TRIG:SEQ1:COUN?
Error Messages
If COUNt >1 when [SENSe[1]]:MRATe <character_data> is set to NORMal or
DOUBle, error –221, "Settings Conflict" occurs.
284
This command sets the number of
triggered events to 10 for the sensor
measurement cycle.
This command queries the number of
triggered events for the sensor
measurement cycle.
U2000 Series Programming Guide