7
SENSe Subsystem
Syntax
Parameters
1
Example
MRAT DOUBle
Reset Condition
On reset, the is set to NORMal.
196
SENS
1
Item
Description/Default
character_data
A numeric value for the measurement :
The default is NORMal.
When a channel is set to NORMal or DOUBle, TRIG:COUNt is set automatically to 1.
Space
:
MRAT
NORMal: 20 readings/second
DOUBle: 40 readings/second
FAST: up to 1000 readings/second
This command sets the sensor to 40
readings/second.
numeric_value
?
Range of Values
1
NORMal
1
DOUBle
FAST
U2000 Series Programming Guide