Agilent Technologies U2000 Series Manual page 34

Usb power sensors
Hide thumbs Also See for U2000 Series:
Table of Contents

Advertisement

1
Power Sensor Remote Operation
power sensor. It may be likely that you do not want to preset these states.
Example
The following example sets the expected power value to –50 dBm and the
resolution setting to 3 using the lower level commands.
ABOR1
CALC1:MATH:EXPR
"(SENS1)"
SENS1:POW:AC:RANG -50
INIT1
FETC1?
16
Aborts current measurement
Displays current measurement
Sets lower range
Causes a measurement
Retrieves the current measurement
U2000 Series Programming Guide