1
Power Sensor Remote Operation
If you send two query commands without reading the response from the first, then attempt
N O T E
to read the second response, you may receive some data from the first response followed
by the complete second response. To avoid this, do not send a query command without
reading the response. When you cannot avoid this situation, send a device clear before
sending the second query command.

Using "*" Commands

Commands starting with a "*" are called common commands. They are
required to perform the identical function for all instruments that are
compliant with the IEEE- 488.2 interface standard. The "*" commands are
used to control reset, self- test, and status operations in the power sensor.

Syntax Conventions

Throughout this guide, the following conventions are used for SCPI command
syntax.
• Square brackets ([]) indicate optional keywords or parameters.
• Braces ({}) enclose one or more parameters that may be included zero or
• Triangle brackets (<>) indicate that you must substitute a value for the
• Bars (|) can be read as "or" and are used to separate alternative parameter

Syntax Diagram Conventions

• Solid lines represent the recommended path.
• Ovals enclose command mnemonics. The command mnemonic must be
54
more times.
enclosed parameter.
options.
entered exactly as shown.
U2000 Series Programming Guide