Power Sensor Remote Operation
1
Naming Frequency Dependent Offset Tables
To rename a frequency dependent offset table use:
MEMory:TABLe:MOVE <string>,<string>
The first <string> parameter identifies the existing table name, and the
second identifies the new table name.
The following rules apply to frequency dependent offset table names:
1 Table names use a maximum of 12 characters.
2 All characters must be upper or lower case alphabetic characters, or
numeric (0- 9), or an underscore (_).
No spaces are allowed in the name.
Reviewing Table Data
To review the data stored in a frequency dependent offset table, use the
following commands:
MEMory:TABLe:SELect "Offset1"
Select the frequency dependent offset table named "Offset1".
MEMory:TABLe:SELect?
Query command which returns the name of the currently selected table.
MEMory:TABLe:FREQuency:POINTs?
Query command which returns the number of stored frequency points.
MEMory:TABLe:FREQuency?
Query command which returns the frequencies stored in the frequency
dependent offset table (in Hz).
MEMory:TABLe:GAIN[:MAGNitude]:POINTs?
Query command which returns the number of offset factor points stored in
the frequency dependent offset table.
MEMory:TABLe:GAIN[:MAGNitude]?
Query command which returns the offset factors stored in the frequency
dependent offset table.
Modifying Data
If you need to modify the frequency and offset factor data stored in a
frequency dependent offset table you need to resend the complete data lists.
If you have retained the original data in a program, edit the program and
resend the data.