Agilent Technologies U2000 Series Manual page 299

Usb power sensors
Hide thumbs Also See for U2000 Series:
Table of Contents

Advertisement

Parameters
Item
Description/Default
source
Available trigger sources:
BUS: the trigger source is a *TRG common command or the
TRIGGER:IMMediate SCPI command.
EXTernal: the trigger source is the trigger input in the sensor.
HOLD: triggering is suspended. The only way to trigger the power sensor is
to use TRIGger:IMMediate.
IMMediate: the trigger system is always true. If INITiate:CONTinuous
is ON the power sensor is continually triggering free (free run mode). If an
INITiate:IMMediate command is sent a measurement is triggered then
the power sensor returns to the idle state.
The trigger source is set to IMMediate on instrument power-up.
N O T E
The MEASure and CONFigure commands automatically set the trigger source to
IMMediate.
The READ? or MEASure commands should not be used if the trigger source is set to HOLD.
Example
TRIG:SOUR IMM
Reset Condition
On reset, the trigger source is set to IMMediate.
U2000 Series Programming Guide
This command configures the sensor for
immediate triggering.
TRIGger Subsystem
Range of Values
EXTernal
HOLD
IMMediate
BUS
279
11