1
Power Sensor Remote Operation
Table 1-5 Bit Definitions - Questionable Status Registers
The condition bits are set and cleared under the following conditions:
Table 1-6 Bit change conditions for Questionable Status Register
44
Bit
Decimal
Number
Weight
0 to 2
-
3
8
4 to 7
-
8
256
9
512
10 to 14
-
15
-
Bit
Meaning
Number
3
POWer
Summary
Definition
Not used
POWer Summary
Not used
CALibration Summary
Power On Self Test
Not Used
Not used (always 0)
EVENts Causing Bit Changes
This is a summary bit for the Questionable POWer Register.
•
SET:
Error –230, "Data corrupt or stale"
•
CLEARED: When no errors are detected by the power
sensor during a measurement covering the causes given
for it to set.
U2000 Series Programming Guide