MEMory Subsystem
MEMory:TABLe:FREQuency
<numeric_value>{,<numeric_value>}
This command is used to enter frequency data into the current selected table.
Any previous frequency list is cleared before the new frequency list is stored.
The frequencies must be entered in ascending order. Entries in the frequency
lists correspond as shown in

Table 6-1 Frequency and Offset Factor List

Ensure that the frequency points you use cover the frequency range of the
signals that you want to measure. If you measure a signal with a frequency
outside the frequency range defined in the table, then the power sensor uses
the highest or lowest point in the table to calculate the calibration
factor/offset.
Depending on available memory, the power sensor is capable of storing 10
frequency dependent offset tables, each containing 80 points.
148