1
Power Sensor Remote Operation
Status Reporting
Status reporting is used to monitor the power sensor to determine when
events have occurred. Status reporting is accomplished by configuring and
reading status registers.
The power sensor has the following main registers:
• Status Register
• Standard Event Register
• Operation Status Register
• Questionable Status Register
• Device Status Register
There are other registers that exist "behind" the main registers, and are
described later in this chapter.
Status and Standard Event registers are read using the IEEE- 488.2 common
commands.
Operation and Questionable Status registers are read using the SCPI STATus
command subsystem.
The General Status Register Model
The generalized status register model shown in
block of the SCPI status system. This model consists of a condition register, a
transition filter, an event register and an enable register. A set of these
registers is called a status group.
36
Figure 1- 6
is the building
U2000 Series Programming Guide