Error Message List
–101
–102
–103
–105
–108
–109
–112
–113
U2000 Series Programming Guide
Invalid character
An invalid character was found in the command string. You may have inserted a
character such as #, $, or % in the command header or within a parameter.
For example, LIM:LOW O#.
Syntax error
Invalid syntax was found in the command string.
For example, LIM:CLE:AUTO, 1 or LIM:CLE: AUTO 1.
Invalid separator
An invalid separator was found in the command string. You may have used a comma
instead of a colon, semicolon, or blank space; or you may have used a blank space
instead of a comma.
For example, OUTP:ROSC,1.
GET not allowed
A Group Execute Trigger (GET) is not allowed within a command string.
Parameter not allowed
More parameters were received than expected for the command. You may have
entered an extra parameter, or added a parameter to a command that does not
accept a parameter.
For example, CAL 10.
Missing parameter
Fewer parameters were received than expected for the command. You omitted one
or more parameters that are required for this command.
For example, AVER:COUN.
Program mnemonic too long
A command header was received which contained more than the maximum 12
characters allowed.
For example, SENSeAVERageCOUNt 8.
Undefined header
A command was received that is not valid for this power sensor. You may have
misspelled the command, it may not be a valid command or you may have the wrong
interface selected. If you are using the short form of the command, remember that it
may contain up to four letters.
For example, TRIG:SOUR IMM.
SYSTem Subsystem